Product Details:
Payment & Shipping Terms:
|
Low Temp. Chamber: | -55ºC~-10ºC | High Temp. Chamber: | +60ºC~+200ºC |
---|---|---|---|
Interior Material: | SUS304 | Exterior Material: | Baked Painting Steel Or SUS304 |
Insulation Material: | Rigid Polyurethane Foam+glass Fiber Wool | Ambient Temp: | +5ºC~+35ºC |
High Light: | temperature shock test chamber,environmental testing chambers |
2-zone stainless steel 500L high low ambient temp thermal shock test chamber
Description
only two chambers: high temp. Chamber and low temp. chamber
Advantage: short conversion time, smaller space needed
Disadvantage: test sample moving. test samples’ size is limited by the basket size
Features of Different Cooling Mode for thermal shock chambers
Since the cold chamber need huge cooling capacity for shocking, the chamber’s cooling mode is often
designed as air-cooled or water-cooled.
Air-cooled type need installation environment:
The room must be equipped with exhauster and air conditioner to make sure heat dissipation timely.
It’s applied to small power products and cool areas.
Water-cooled type need installation conditions:
The customer must prepare water pipe and water tower (located outside) accordingly.
The water must be pure, or it will block the pipes and result in some maintenance trouble with time going.
It’s often applied to large system with the features of better heat dissipation, longer service time, low
noise but slightly higher price if count the fee of water tower and water pipes.
Product Specifications
Interior size(W*H*D)mm | 810*710*860 |
Exterior size(W*H*D)mm | 710*610*630 |
Low Temp. chamber | -55ºC~-10ºC |
Pre-cooling time | +20ºC~-55ºC, Within 60 min |
Test. chamber | -40ºC~+150ºC |
High Temp. chamber | +60ºC~+200ºC |
Pre-heating time | +60ºC~+200ºC ,Within 25 min |
Recovery time | High-Temp Exposure 30 min |
Low-Temp Exposure 30 min | |
Recovery time within 3 min | |
Safety device | NFB(No fuse breaker), over pressure, over heat and over current protection for compressor,over temperature protection, over load protection for blower, dry heat protection |
Standard accessory | Cable port(portΦ50mm)*1, shelves*2 |
Interior material | Stainless steel plate(SUS 304) |
Exterior material | Baked painting steel or stainless steel(SUS304) |
Insulation material | Rigid polyurethane Foam+glass fiber wool |
Refrigeration system | Water Cooled |
Semi-hermetic compressor, cascade refrigeration system, CFC free refrigerant(HFC-507 and HFC-23) | |
Controller | Panel:7-inch LCD Touch panel, Chinese or English display selectable |
Operation model:Program or fix point running | |
Memory capacity:120programs,1200steps, all repeat 999cycles,part repeat 99 cycles. | |
Accuracy:Temp.:0.1% of F.S±1 digit Humidity:0.1% of F.S±1 digit | |
Input:Temp.:pt-100 Humidity:pt-100 or ~5V DCV | |
Ambient Temp. | +5ºC~+35ºC |
Power | AC 380±10% 50Hz ,3 phase 4 wires +Ground Wires |
Note:1. We reserve the right to change specifications without prior notice 2. Customized sizes and configurations available | |
2. Customized sizes and configurations available |
Product Standards Implement and Met
GB/T 2443.1-2001 Test A: Low temp. Test;
GB/T 2443.2-2001 Test B: Low temp. Test;
GB/T 2443.22-2002 Test N: Low temp. Test;
GJB 150.3-1986 High Temp. Test;
GJB 150.4-1986 Low temp. Test;
GJB 150.5-1986 Temp. Impact Test;
(Load refer to 5.5.5 no active load)
Product details
Product Packaging
Each set protected with resin fiber and PP film, then put into strong wooden
case with operation mannual and video inside.
Product After-sale service
If you face any problems in future during testing process we promise we will
give you a reply within 48 hours and solution provide within 3 working days. We
can offer operation video and English operation muanual. We can provide
video-conference. If customers require on-site service, the transport and travel
expense shall be borne by customer.
Contact Person: Anna Hu
Tel: +8618098282716