Product Details:
Payment & Shipping Terms:
|
Pre-set Function: | Boot Time Can Be Set Freely And Machine Runs Automatically When Turning On Power | Humidity Uniformity: | 3.0%RH (no-load) |
---|---|---|---|
Interior Material: | SUS 304 | Heat Up Rate: | From -70℃ ↑ +180℃ Average 20℃/min No Load .(According To IEC60068-3 Standard ) |
Humidity Deviation: | ±3.0%RH(>75%RH) ±5.0%RH(≤75%RH) | Packing: | Rigid Foam And Wooden Box |
High Light: | climatic test chambers,Environmental Test Chamber |
Able to accurately simulate a wide range of complicated natural environments, and is suitable for reliability
test in industrial products. Meet GB5170.2.3.5.6-95 standard requirements of environmental testing
equipment and test methods for the basic parameters of electric and electronic products under the condition
of humidity, low temperature, high temperature, and constant heat.
It is suitable for tests the products of electrical, electronic, mechanical and other products, parts and
materials under the conditions of high and low temperature and high temperature and humidity. For quality
and reliability test of simulating temperature and humidity changes in electrical, electronic, mechanical and
other products, parts and materials.
Specification
Capacity(L) |
64L
|
Temperature range |
L :-40℃ ~ +150℃ R : :-20℃ ~ +150℃ S:-70℃ ~ +150℃ |
Temp Deviation |
≦ ± 2.0℃ |
Size |
Workspace volume: W 400 × H 400 × D 400 mm
Exterior size: W 600 × H 1550 × D 1220 mm (about) |
Temp Fluctuation |
±0.5℃ |
Heat up rate |
Heating up rate: -40℃↑ +100℃ about 45min (no-load, ambient temperature +25℃)
Cooling rate: 20 ℃to -40℃about 60min (no-load, ambient temperature +25℃) |
Humidity range |
20%R.H~98%R.H
|
Door |
Single door, left open. Heating wire was installed at the door frames to prevent
condensation at low temperatures. |
Humidity uniformity |
3.0%RH (no-load) |
Load |
Product calorific value 1000w
|
Humidifier |
Water basin heating and humidification method;
Stainless steel sheathed heater;
Heater control: no-contact control (SSR);
Water level control device, heater anti-dry device |
Electrical Control System
Control cabinet |
A. Emergency stop switch
B. Power switch
C. Over-temperature protection
D.RS-485 interface |
Protection System |
A. Heater protection switch if no water
B. Humidifier protection switch if no water
C. Heater over-current circuit breaker
D. Humidifier over-current circuit breaker
|
Protection System |
E. Circulating fan over-current overload protection
F. Compressor high voltage protection switch
G. Compressor overheat protection switch
H. Compressor over-current protection switch
I. Over-voltage under-phase protection switch
J. Circuit Breakers
K. Leakage switch
L Low humidifier protection
M. Water tank low water level warning
Controller noise isolation protection
N. Zero-crossing gate fluid power controller |
Alarm indicator |
Equipment stops running and sends audible alarm when the above protection appears,
meanwhile, fault, causes and solutions will be displayed on the screen. |
Ambient temp. and humidity |
5 ~ 35℃ |
Power |
AC 3ψ4W 380V 50Hz (R, S, T, N plus ground) (voltage fluctuation ≦ ± 10%)
|
Grounding |
Grounding resistance≦4Ω
|
Function |
Fault alarm and cause handling prompts, power failure protection, the temperature upper and lower limit protection, timer function (automatic start and automatic stop running), self-diagnostic function. |
Network Connection |
Can be connected to Ethernet via professional software, Remote control & assistance function and data collection can be achieved through network, Multi machine can be controlled simultaneously |
Software environment |
Windows XP or Windows 7 / WIN8 operating syste |
Feathers
● GB-2423.1-2008 (IEC68-2-1)Test A: Low Temperature Test
● GB-2423.2-2008(IEC68-2-2)Test B: High Temperature Test
● GJB360.8-2009(MIL-STD.202F) High Temperature Life Test
● GJBl50.3-2009(MIL-STD-810D) High Temperature Test
● GJBl50.4-2009 (MIL-STD-810D) Low Temperature Test
● GB2423.3-2008(IEC68-2-3)Test Ca: Constant Heat Test
● GB2423.4-2008(IEC68-2—30) Test Db: Damp Heat Alternative Test
● GJBl50.9-2009 (MIL-STD-810D) Damp Heat Test
For more detail information ,kindly contact : anna@komegtech.com whats app ,wechat :+86-18029096856
Photos show
Contact Person: Anna Hu
Tel: +8618098282716